TopPette Single-channel Fixed Volume Mechanical Pipette

Ergonomic design, lightweight plunger force can reduce repetitive strain injury. The digital display can clearly read the volume setting. Tip ejector for easy and fast needle tip replacement.
Each pipette is calibrated in accordance with ISO8655 and each micro pipette is supplied with an individual test certificate.

Inquiry
(+86)010-5653 8958
  • Product description
  • Product parameter
  • Product video

TopPette Single-channel Fixed Volume 

Mechanical Pipette

TopPette single-channel.内容3.jpg

Features:

* Lightweight, ergonomic, low force design

* Digital display clearly reads volume setting
* The pipettes cover volume range of 0.1μl to 10ml
* Easy to calibrate and maintain with tool supplied
* Design helps avoid repetitive strain injuries
* Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
* The low part is available for autoclaving

Application:

Best suited for clinical diagnostics and control analysis labs.
Best suited for molecular biology, microbiology, immunology, cell culture, analytical chemistry, biochemistry and genetic labs.
Best suited for ELISA & CLIA diagostics test, molecular screening, kinetic studies, DNA application etc.


Fixed Volume Pipettes
Volume RangeIncrementTest VolumeAccuracy errorPrecision error



%μl%μl
5μl-5μl1.30%0.0651.20%0.06
10μl-10μl0.80%0.080.80%0.08
20μl-20μl0.60%0.120.50%0.1
25μl-25μl0.50%0.1250.30%0.075
50μl-50μl0.50%0.250.30%0.15
100μl-100μl0.50%0.50.30%0.3
200μl-200μl0.40%0.80.20%0.4
250μl--250μl0.40%10.20%0.5
500μl-500μl0.30%1.50.20%1
1000μl-1000μl0.30%30.20%2
2000μl-2000μl0.30%60.15%3
5000μl-5000μl0.30%150.15%7.5