On board software with large screen panel, easy for use:7-inch color touch screen shows all parameters, test results and real time testing curve.
Portable and light : the weight is only 4kg. It can be used with vehicle power supply for immediate on-site detection.
Sensitive and efficient optical signal detection system, multi-channel fluorescence without crosstalk: based on unique Fresnel lens, high-efficiency PMT and maintenance-free LED. 16-hole scanning takes only 1s.
Ultra-fast temperature control system: unique liquid circulation refrigeration.
Compatible with 0.2mL PCR transparent tubes or eight stripes, can measure 16 samples at a time.
Software functions can be customized.
4G data memory. Testing results can be exported with USB or printed.
Accurate 16
On board software with large screen panel, easy for use:7-inch color touch screen shows all parameters, test results and real time testing curve.
Portable and light : the weight is only 4kg. It can be used with vehicle power supply for immediate on-site detection.
Sensitive and efficient optical signal detection system, multi-channel fluorescence without crosstalk: based on unique Fresnel lens, high-efficiency PMT and maintenance-free LED. 16-hole scanning takes only 1s.
Ultra-fast temperature control system: unique liquid circulation refrigeration.
Compatible with 0.2mL PCR transparent tubes or eight stripes, can measure 16 samples at a time.
Software functions can be customized.
4G data memory. Testing results can be exported with USB or printed.
Sample capacity | 2×8 0.2ml tube/strips,clear |
Reaction volume | 10-50μl |
Touch screen | YES |
Thermal cycle technology | Peltier, liquid circulation refrigeration |
Max. Heating/Cooling rate | 6.0°C/s |
Heating temperature range | 4 – 100 °C |
Temperature accuracy | ±0.2°C |
Temperature uniformity | ±0.2℃ @60℃ ±0.3℃ @95℃ |
Excitation light source | Monochrome high efficiency LED |
Detection device | PMT |
Detection mode | Fast hole-by-hole scanning |
Fluorescent channels | Standard:FAM/ SYBR, VIC/HEX/ TET Optional:ROX,CY5/TAMRA |
Sensitivity | Single copy gene |
Dynamic range | 10 orders of magnitude copies |
Analysis mode | CT value analysis, absolute quantitative analysis |
Dimension [W×D×H] | 205×190×98mm |
Power Supply | 100-240V or Optional 12V, 10A mobile power |
Weight | 4kg |